Model Number 22-4035 |
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/27/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that, during a shoulder arthroscopy using the 'firstpass suture passer', the trap door came off when loading the needle on back table.Used a second needle and trapdoor kit and was successfully loaded on back table.When surgeon went to pass suture through the rotator cuff, the trap door came dislodged.The free floating trap door was removed using a cuff grasper.The procedure was completed with a non-significant delay using a smith and nephew backup device.No further complications were reported.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).H10 h3, h6: the reported device was received for evaluation.There was no relationship found between the device and the reported event.A visual evaluation showed the device did not have a suture capture or needle when returned.No deficiencies are visible.A functional evaluation showed the test needle and suture capture were loaded into the device.Pulling the trigger to the first stop closed and locked the jaw.Pulling the trigger to the second stop, deployed the needle, which caught the test suture and captured it.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Based on the information provided, the free-floating trap door was removed from inside of the patient with a cuff grasper.Since there were no other complications reported, no further clinical/medical assessment is warranted at this time.Should any additional relevant patient information be provided, this case would be re-assessed.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
|
Event Description
|
It was reported that, during a shoulder arthroscopy using the 'firstpass suture passer', the trap door came off when loading the needle on back table.Used a second needle and trapdoor kit and was successfully loaded on back table.When surgeon went to pass suture through the rotator cuff, the trap door came dislodged.The free floating trap door was removed using a cuff grasper.The procedure was completed with a non-significant delay using a smith and nephew backup device.No further complications were reported.Patient current status is fine.
|
|
Search Alerts/Recalls
|
|