Model Number 45007 |
Device Problem
Suction Problem (2170)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/25/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that a loss of aspiration occurred.The target lesion was located in the tibial artery.A 2.1mm jetstream was selected for use.During the procedure, a loss of performance and aspiration occurred in blades down mode.There were no error messages or resistance noted.Another of the same device was used to complete the procedure.There were no patient complications.
|
|
Event Description
|
It was reported that a loss of aspiration occurred.The target lesion was located in the tibial artery.A 2.1mm jetstream was selected for use.During the procedure, a loss of performance and aspiration occurred in blades down mode.There were no error messages or resistance noted.Another of the same device was used to complete the procedure.There were no patient complications.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: returned product consisted of a jetstream xc-2.1 atherectomy catheter.The device was visually examined for any shaft damage and the functional testing of the device was completed.The device showed shaft damage in the form of a severe kink located 37.3cm from the tip.There was also a small kink located 76.5cm from the tip.Multiple small bends were also noticed on the shaft.Functional analysis was done by completing the setup procedure per the instructions for use (ifu).The devices blades did not rotate.Aspiration testing of the device was done per the test procedure.The device is tested by using a 100ml beaker of water.The devices tip is submerged in a beaker of fluid and the device is run for a period of 1 minute.The final milliliters of fluid that the beaker shows after the 1minute run time is subtracted from the starting milliliters and the final number is the total that was aspirated (100ml-3ml=97ml).Test results showed that this device did not perform as designed per the test procedure specification sheet withdrawing 3ml of fluid in the 1minute time frame.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
|
Search Alerts/Recalls
|