OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problem Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2022
Event Type  malfunction  
Manufacturer Narrative
The suspect device has been returned to olympus for evaluation, the service center observed did not confirm the reported event as the lamp illuminated properly at maximum output.It was noted a non-olympus lamp was installed in the unit.Additionally, the lamp was not seated properly into the heat sink and the lamp bolts/washers were mis-aligned.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus that during preparation for a procedure the evis exera iii xenon light source illumination is low when turned in the max position and is not bright.There was no patient death, harm, or injuries reported with this event.
 
Manufacturer Narrative
This supplemental report is to inform that upon further review, this is not a reportable malfunction.Per the legal manufacturer, there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13560616
MDR Text Key296387815
Report Number8010047-2022-03210
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/25/2022
Initial Date FDA Received02/18/2022
Supplement Dates Manufacturer Received04/15/2022
Supplement Dates FDA Received04/18/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-