Model Number 595000-001 |
Device Problems
No Apparent Adverse Event (3189); Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The freedom driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4) initial.
|
|
Event Description
|
The customer, a syncardia authorized distributor, reported that the freedom driver exhibited a fault alarm while supporting a patient.There was no reported adverse patient impact.The customer also reported that the patient was subsequently changed to a backup driver.
|
|
Manufacturer Narrative
|
The customer-reported fault alarm was confirmed by examining the driver's alarm history which revealed a 0x1e fault code, which indicates mpu regulator 1 too low.The driver passed all functional testing, including a 4- hour observation run.Since the "1e" fault code can record during a battery exchange, a battery exchange test was performed with no alarms.The root cause of the customer-reported fault alarm could not be conclusively determined.The freedom driver performed as intended with no evidence of a device malfunction.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
|
|