• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL SALINE SYRINGES; SALINE, VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL SALINE SYRINGES; SALINE, VASCULAR ACCESS FLUSH Back to Search Results
Model Number 306546
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2022
Event Type  malfunction  
Manufacturer Narrative
It was reported the plunger of the syringe was hard to push.As a sample was not returned, a thorough sample investigation could not be completed.It could be possible that the customer is getting some products that are towards the high specification limit and are related to the symptom reported by the customer since they require extra force than normal to expel the solution.A device history record review was completed for provided material number 306546, lot 1167771.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.Previous investigations have revealed that improper silicone application within the syringe barrel can create plunger resistance.Several quality initiatives have been implemented on our manufacturing line to ensure that the silicone application is properly applied during the manufacturing process.Additionally, a monitoring program is also in place to verify the silicone is uniformly applied to the syringe barrel.Based on the investigation and with no sample analysis the symptom reported by the customer could not be confirmed.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported while using bd posiflush¿ normal saline syringes the plunger was difficult to move.There was no report of patient impact.The following information was provided by the initial reporter: the customer complained that the plunger of the syirnge was hard to push.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NORMAL SALINE SYRINGES
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13580390
MDR Text Key286775849
Report Number1911916-2022-00091
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public30382903065463
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number306546
Device Catalogue Number306546
Device Lot Number1167771
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-