Model Number P1190A000067 |
Device Problem
Unintended Movement (3026)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The hillrom technician found the siderail cable assembly needed to be replaced.Per the hillrom service manual the advanta¿ 2 should be subject to an effective maintenance program.Test each of the buttons to check that they activate the correct function and that they do not work intermittently by pressing each button for several seconds.Each movement must be continuous.Replace the pendant if necessary.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the siderail cable to resolve the issue.Based on this information, no further action is required.
|
|
Event Description
|
Hillrom received a report from a hillrom technician stating that a button was stuck, making the head function operate when plugged into ac power.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
|
|