• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE DISTAL COVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE DISTAL COVER Back to Search Results
Model Number MAJ-2315
Device Problems Entrapment of Device (1212); Detachment of Device or Device Component (2907)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Type  Injury  
Event Description
The customer reports during two separate (unspecified) exams using an evis exera iii duodenovideoscope with a single use distal cover, the distal cover fell off the scope (and into the patient).During procedure 1 of 2, the distal cover could not be located to remove it from the patient.During procedure 2 of 2, the distal cover was found and retrieved from near the bite block.No additional consequences to the patients were reported as a result of these occurrences.Additional details regarding the patients and reported events were requested.The customer stated that the physician could not remember in which patient cases this occurred, and therefore no further patient event could be provided.The nurse assisting in both cases with this physician stated ¿i think it could have been my fault for not making sure the cap was secure on the tip.I now push it on and then pull on it to make sure it will not fall off.I have not had any further problems.Case with patient identifier (b)(6) reports the single use distal cover used in procedure 1 of 2 (which was not retrieved).Case with patient identifier (b)(6) reports the tjf-q190v used in procedure 1 of 2.Case with patient identifier (b)(6) reports the single use distal cover used in procedure 2 of 2 (which was retrieved).Case with patient identifier (b)(6) reports the tjf-q190v used in procedure 2 of 2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE DISTAL COVER
Type of Device
DISTAL COVER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key13601103
MDR Text Key286165175
Report Number2951238-2022-00323
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170403019
UDI-Public04953170403019
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/28/2022,02/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAJ-2315
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/28/2022
Date Report to Manufacturer01/28/2022
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
TJF-Q190V
Patient Outcome(s) Other;
-
-