NULL PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C37101329-NLJ |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during the pre-use check, leakage of air from the breathing circuit was observed.No patient involvement.No additional information is available for this complaint.
|
|
Manufacturer Narrative
|
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No problems or issues were identified during this device history record review.A product sample was received for evaluation.Visual and functional testing were performed.The sample was visually inspected: a hole was found due to embedded resin in the circuit.The device presented leakage.The root cause of the reported issue was found to be undetermined.A corrective and preventative action has been opened to address the reported issue.
|
|