Model Number CI-1600-04 |
Device Problems
Mechanical Problem (1384); Output below Specifications (3004); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 02/14/2022 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing decreased performance.Device testing revealed results within normal limits.Programming adjustments were made, however, the issue did not resolve.The recipient device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the electrode array was severed, as well as tool damage to the top cover.These anomalies are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the hires ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|