• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® GOODWIN SOUND; DILATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® GOODWIN SOUND; DILATION CATHETER Back to Search Results
Model Number 042828
Device Problem Corroded (1131)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/04/2022
Event Type  malfunction  
Event Description
It was reported that the customer received a partial order of quite a few goodwin sound replacements as it consistently had issues with rust and discoloration in the lumen.Also stated that upon receiving the new replacements, they were inspected with a borescope and discoloration was noted prior to any reprocessing.Per notification from investigator on (b)(6) 2022 based on sample returned, it was reported that the dilation catheters were corroded.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.The device did not fail to meet relevant specifications.As the product was not used on a patient it was not used for diagnostic or treatment purposes.The product had not caused the reported failure.No root cause could be found because the reported event was unconfirmed.Two, 28 fr sound dilator were returned.No rusting or discoloration was noted.Device meets specifications per the visual requirements which states "surface must have a mirror bright finish".The lot number was unknown; therefore, the device history record could not be reviewed.Labeling review is not required as the reported event is unconfirmed.The actual/suspected device was inspected.
 
Event Description
It was reported that the customer received a partial order of quite a few goodwin sound replacements as it consistently had issues with rust and discoloration in the lumen.Also stated that upon receiving the new replacements, they were inspected with a borescope and discoloration was noted prior to any reprocessing.Per notification from investigator on 03feb2022 based on sample returned, the additional records were created based on the material numbers of the samples received.
 
Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.The device did not fail to meet relevant specifications.As the product was not used on a patient it was not used for diagnostic or treatment purposes.The product had not caused the reported failure.No root cause could be found because the reported event was unconfirmed.Two, 28 fr sound dilator were returned.No rusting or discoloration was noted.Device meets specifications per the visual requirements which states "surface must have a mirror bright finish".The lot number was unknown; therefore, the device history record could not be reviewed.Labeling review is not required as the reported event is unconfirmed.The actual/suspected device was inspected.
 
Event Description
It was reported that the customer received a partial order of quite a few goodwin sound replacements as it consistently had issues with rust and discoloration in the lumen.Also stated that upon receiving the new replacements, they were inspected with a borescope and discoloration was noted prior to any reprocessing.Per notification from investigator on 04feb2022 based on sample returned, the additional records were created based on the material numbers of the samples received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® GOODWIN SOUND
Type of Device
DILATION CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13636011
MDR Text Key287028217
Report Number1018233-2022-00954
Device Sequence Number1
Product Code FBX
UDI-Device Identifier00801741047305
UDI-Public(01)00801741047305
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number042828
Device Catalogue Number042828
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/04/2022
Initial Date FDA Received03/01/2022
Supplement Dates Manufacturer Received04/20/2022
04/20/2022
Supplement Dates FDA Received05/13/2022
05/16/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-