• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAPTIVATOR EMR; ENDOSCOPE AND/OR ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAPTIVATOR EMR; ENDOSCOPE AND/OR ACCESSORIES Back to Search Results
Model Number M00561610
Device Problems Use of Device Problem (1670); Suction Problem (2170)
Patient Problem Perforation (2001)
Event Date 01/28/2022
Event Type  Injury  
Manufacturer Narrative
The complainant was unable to provide the lot number.Therefore, the manufacture and expiration dates are unknown.(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a captivator endoscopic mucosal resection device was used during a endoscopic mucosal resection (emr) procedure performed on (b)(6), 2022.When the device was opened the blue cap was dropped/lost so the blue cap was missing.The technician used their finger to complete suction loop instead.The device was successfully used in the antrum.When used in the fundus, the band captured muscularis due to too much suction and the band was deployed over the whole wall and was not realized until after it was cut, and a perforation occurred.The procedure was successfully completed with a non-bsc clip device.The patient outcome was reported to be fully recovered.There was no additional intervention required and the patient was not admitted to the hospital beyond standard of care.The patient condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIVATOR EMR
Type of Device
ENDOSCOPE AND/OR ACCESSORIES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13639856
MDR Text Key287425499
Report Number3005099803-2022-00986
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729842699
UDI-Public08714729842699
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140726
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00561610
Device Catalogue Number50661
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/31/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-