• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3367-40QC
Device Problem Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2022
Event Type  malfunction  
Event Description
It was reported the asymptomatic patient presented in clinic.During procedure on (b)(6) 2022, it was observed the setscrew of the implantable cardiac defibrillator (icd) was stuck and couldn't be disconnected.The icd was replaced during procedure.The patient was stable.
 
Manufacturer Narrative
The reported event of set screw anomaly was confirmed.The device was above elective replacement indicator (eri) upon receipt.Analysis revealed the set screw was stripped and clogged with material.The set screw anomaly was consistent with having occurred during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13644406
MDR Text Key286439029
Report Number2017865-2022-04263
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeEC
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model NumberCD3367-40QC
Device Lot NumberP000131662
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/10/2022
Initial Date FDA Received03/01/2022
Supplement Dates Manufacturer Received05/06/2022
Supplement Dates FDA Received05/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-