• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® GOODWIN SOUND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® GOODWIN SOUND Back to Search Results
Model Number 042830
Device Problem Corroded (1131)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/15/2021
Event Type  malfunction  
Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.The device did not fail to meet relevant specifications.As the product was not used on a patient it was not used for diagnostic or treatment purposes.The product had not caused the reported failure.No root cause could be found because the reported event was unconfirmed.14 fr sound dilator was returned.No rusting or discoloration was noted.Device meets specifications per the visual requirements which states "surface must have a mirror bright finish".The lot number was unknown; therefore, the device history record could not be reviewed.Labeling review is not required as the reported event is unconfirmed.The actual/suspected device was inspected.
 
Event Description
It was reported that the customer received a partial order of quite a few goodwin sound replacements as it had consistently had issues with rust 042822 and 042820 and discoloration 042816 and 042814 in the lumen.It was also stated upon receiving the new replacements, they were inspected with a borescope and discoloration was noted prior to any reprocessing.Per follow up via email on 16aug2021, the lot numbers of the affected sounds which are mixed with new box and never processed.The new box, which was processed initially, never used on patient.This was the ongoing issue for the customer and consistently replacing these.The customer checked and verified internal process and follow the ifu verbatim, flushing with sterile water at point of use.The box and packaging that customer received was for 11 sounds.They have total of 24 sounds 042824 and 042826.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® GOODWIN SOUND
Type of Device
BARD® GOODWIN SOUND
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13652550
MDR Text Key287027568
Report Number1018233-2022-00980
Device Sequence Number1
Product Code FBX
UDI-Device Identifier00801741047312
UDI-Public(01)00801741047312
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number042830
Device Catalogue Number042814
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/04/2022
Initial Date FDA Received03/02/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-