NULL PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number CA13E0/4K1/004JP |
Device Problems
Leak/Splash (1354); Material Puncture/Hole (1504)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during the pre-use check, leakage of air from the product was observed.A further check confirmed a pinhole in the breathing bag.No patient involvement.No additional information is available for this complaint.
|
|
Manufacturer Narrative
|
One unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.The problem source is manufacturing.Dhr review was done, no issues related to the original complaint were found.
|
|