MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Failure to Deliver Energy (1211); Battery Problem (2885); Charging Problem (2892); Connection Problem (2900); Material Integrity Problem (2978)
|
Patient Problems
Dysphasia (2195); Cognitive Changes (2551); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/28/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other relevant device(s) are: product id: 37761, serial#: (b)(4), product type: recharger.Other relevant device(s) are: product id: 37761, serial/lot #: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the desktop charger connector pin pulled off the wire leaving the red wire exposed on (b)(6) 2022.Additional information received from the consumer reported without the desktop charger to charge the recharger they couldn¿t charge the implant.The patient was showing signs of symptom return (having trouble talking/understanding things).The programmer was used to check the implant and the implant was off due to the battery being depleted.Unrelated programmer issue event captured in (b)(4).
|
|
Manufacturer Narrative
|
Continuation of d10: product id 37761 lot# serial# (b)(6): product type recharger h3: analysis of the desktop charger c0627430 revealed that the cable assembly was frayed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|