WRIGHT MEDICAL TECHNOLOGY INC INFINITY TIBIAL TRAY SIZE 5 TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 33650005
Device Problem Loss of Osseointegration (2408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2022
Event Type  Injury  
Event Description
It was reported that the patient underwent a total ankle replacement due to tibial tray loosening.
 
Manufacturer Narrative
Device not available for evaluation as it remains implanted in the patient.If additional information becomes available, it will be provided on a supplemental report.
 
Manufacturer Narrative
Please note the correction in section d4 lot# more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.There are many clinical factors that can affect the results of any surgery, such as surgical technique, pre-operative and post-operative care, the implant, patient pathology and daily activity.Medical profession reviewed the images and noted: there are indeed signs of cyst formation (anteriorly) and some osteolysis between the tibial tray and the tibia in the posterior aspect.A loosening cannot be confirmed from the provided ct images.That diagnosis can only be made with a significant position change of the implant, or as in this case, comparison with the early postoperative imaging.That would be helpful to see if the bone changes in this case also denote loosening.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event Description
It was reported that the patient underwent a total ankle replacement due to tibial tray loosening.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY TIBIAL TRAY SIZE 5 TOTAL ANKLE SYSTEM
Type of Device
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13664215
MDR Text Key289078273
Report Number3010667733-2022-00114
Device Sequence Number1
Product Code HSN
UDI-Device Identifier00840420122520
UDI-Public00840420122520
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K140749
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number33650005
Device Catalogue Number33650005
Device Lot Number1687347
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/08/2022
Initial Date FDA Received03/03/2022
Supplement Dates Manufacturer Received03/31/2022
Supplement Dates FDA Received04/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age59 YR