ROCHE DIAGNOSTICS TSH ELECSYS COBAS E 200 V2; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Model Number TSH
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2022
Event Type  malfunction  
Event Description
The initial reporter received questionable elecsys ft4 iii, elecsys ft3 iii, and tsh elecsys results for 1 patient on a cobas 8000 cobas e 602 module, serial number, (b)(4).This medwatch covers the alleged results for the tsh assay.Please refer to medwatch (b)(4) for the alleged ft3 assay results and medwatch (b)(4) for the alleged ft4 assay results.The results were reported to the patient's endocrinologist, who requested the sample test be repeated due to the results not matching the patient's clinical picture.The patient's results obtained with cmia (chemiluminescent microparticle immunoassay) technology were considered to be clinically correlating with the patient.The customer suspects an interference.
 
Manufacturer Narrative
Sample material was requested for investigation.The investigation is ongoing.
 
Manufacturer Narrative
The customer's calibration and qc results and sample pre-analytic details were requested but not provided.The sample could not be provided for investigation.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TSH ELECSYS COBAS E 200 V2
Type of Device
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key13666271
MDR Text Key292979556
Report Number1823260-2022-00573
Device Sequence Number1
Product Code JLW
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTSH
Device Catalogue NumberASKU
Device Lot Number574298
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CALCIUM GLUCONATE; INOTROPE; INSULIN INFUSION; IV FLUIDS; MAGNESIUM SULFATE; NORADRENALINE; ORAL HYPOGLYCEMIC; VITAMIN D
Patient Age54 YR