• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. AUTOCAT2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. AUTOCAT2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000320
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/03/2022
Event Type  malfunction  
Event Description
There was no patient involvement.It was reported that the pump alarmed "purge failure." as a result, the pneumatic control system assembly was replaced, and the machine was returned to service.
 
Manufacturer Narrative
(b)(4).There is an investigation ongoing into this reported complaint.A supplement follow-up mdr report will be submitted with the investigation results upon completion of the investigation.
 
Manufacturer Narrative
(b)(4).Teleflex received the device for investigation.The reported complaint of purge failure is confirmed.The returned pcs assembly's balloon pressure transducer was faulty, which caused a purge failure alarm.The root cause of this complaint is undetermined.Corrections have been established for this issue as a result of a non-conformance and this device was manufactured prior to the release of those corrections.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.
 
Event Description
There was no patient involvement.It was reported that the pump alarmed "purge failure." as a result, the pneumatic control system assembly was replaced, and the machine was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT2 WAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kaylia pen
16 elizabeth drive
chelmsford, MA 01824
MDR Report Key13666861
MDR Text Key286567933
Report Number3010532612-2022-00043
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902051715
UDI-Public30801902051715
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K060309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN000320
Device Catalogue NumberIAP-0500I
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/03/2022
Initial Date FDA Received03/04/2022
Supplement Dates Manufacturer Received04/04/2022
Supplement Dates FDA Received04/05/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
NI.; NI.
-
-