• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. MEDEX FILTER SET; FILTER, INFUSION LINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. MEDEX FILTER SET; FILTER, INFUSION LINE Back to Search Results
Model Number MX448HFB
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2022
Event Type  malfunction  
Event Description
It was reported that the iv fluid is not running through the administration set.There was no patient injury reported.
 
Manufacturer Narrative
One filter set was received for evaluation in relation to the reported event.Evaluation of the device is currently in progress.Once the evaluation is complete, a follow-up report with evaluation results will be submitted.
 
Manufacturer Narrative
This mdr was generated under protocol b10010116, as a result of warning letter cms# (b)(4).A product sample was received for evaluation.Visual and functional testing were performed.Sample went through water leak and occlusion test and it did not exhibit leakage or occlusion condition.The root cause of the reported issue cannot be determined, as the returned sample did not exhibited occlusion nor defective condition.Actions were taken to mitigate the reported issue: awareness meeting with operators and quality inspectors about the failure mode reported on this complaint was held.No problems or issues were identified during this device history record review.
 
Manufacturer Narrative
Other text: this mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A product sample was received for evaluation.Visual and functional testing were performed.Sample went through water leak and occlusion test and it did not exhibit leakage or occlusion condition.The root cause of the reported issue cannot be determined, as the returned sample did not exhibited occlusion nor defective condition.Actions were taken to mitigate the reported issue: awareness meeting with operators and quality inspectors about the failure mode reported on this complaint was held.No problems or issues were identified during this device history record review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDEX FILTER SET
Type of Device
FILTER, INFUSION LINE
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6250 shier rings road
dublin OH 43016
Manufacturer (Section G)
NULL
6250 shier rings road
dublin OH 43016
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key13672452
MDR Text Key286772296
Report Number3012307300-2022-04347
Device Sequence Number1
Product Code FPB
UDI-Device Identifier10351688507457
UDI-Public10351688507457
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K860803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMX448HFB
Device Catalogue NumberMX448HFB
Device Lot Number4180739
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/04/2022
Initial Date FDA Received03/04/2022
Supplement Dates Manufacturer Received02/03/2023
02/03/2023
Supplement Dates FDA Received03/15/2023
03/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-