Catalog Number AASLE08080
Device Problems Loss of or Failure to Bond (1068); Positioning Failure (1158); Misfire (2532)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2021
Event Type  malfunction  
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the covera vascular covered stent that are cleared in the us.The pro code and 510 k number for the covera vascular covered stent are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation as well as photos and videos were provided for review.The investigation of the reported event is currently underway.(expiry date: 03/2023).
 
Event Description
It was reported that during a stent placement procedure, the device allegedly failed to deploy.The procedure was completed using another device.There was no reported patient injury.
 
Event Description
It was reported that during a stent placement procedure, the device allegedly failed to deploy.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the covera vascular covered stent that are cleared in the us.The pro code and 510 k number for the covera vascular covered stent are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned to the manufacturer for evaluation.The delivery system was found disassembled upon receipt; the stent was found partially deployed.The slide block was found no longer connected to proximal sheath.Based on the condition of the sample it is concluded that the slide block getting disconnected from the proximal sheath, led to the reported impossibility to deploy the stent completely.A definite root cause is not yet available.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.Based on the instruction for use supplied with this product delivery system specific events that could be associated with clinical complications include but are not limited to failure to deploy and high deployment forces.Regarding correct deployment the instruction for use states "maintain a stationary hold on the white stability sheath during covered stent deployment (¿).Hold the white stability sheath as close as possible to the introducer without touching the dark brown moving catheter of the distal catheter assembly.Maintain the remainder of the white stability sheath (.) relaxed and avoid tension." regarding preparation and accessories, the instruction for use states: "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated." and "0.035 inch (0.89 mm) guidewire of appropriate length (¿) introducer sheath with appropriate inner diameter and length." h10: d4 (expiry date: 03/2023), g3 h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVERA PLUS VASCULAR COVERED STENT
Type of Device
VASCULAR COVERED STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13683971
MDR Text Key287256178
Report Number9681442-2022-00069
Device Sequence Number1
Product Code PFV
UDI-Device Identifier00801741136184
UDI-Public(01)00801741136184
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
P170042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAASLE08080
Device Lot NumberANFQ2026
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/04/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-