SMITH & NEPHEW, INC. GII QUICK CONNECT HANDLE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71440044 |
Device Problems
Break (1069); Positioning Failure (1158)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that a genesis ii quick connect handle/alignment tower was found broken.The button fails to engage.As this was noticed upon inspection, there was no patient involvement.
|
|
Manufacturer Narrative
|
The device was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|