MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
|
Back to Search Results |
|
Model Number BI70002000 |
Device Problem
Unintended Movement (3026)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/04/2022 |
Event Type
malfunction
|
Event Description
|
Medtronic received information regarding a navigation device being used for a ortho procedure.It was reported that when trying to put the break on the imaging system it would not break.No impact on patient outcome.There was no delay.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information, regarding an imaging device.
|
|
Manufacturer Narrative
|
H3: a medtronic representative, went to the site to test the equipment.Testing revealed, that the clutch was partially engaged.The drive motors was fully engaged and the system works as intended.The imaging system, then passed the system checkout and was found to be fully functional.B01, c07, d02 applicable codes.Concomitant medical products: other relevant device(s) are: product id: bi71000174, sn/lot#: unknown.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from a manufacturer representative indicated there was no patient present.The reported issue occurred during transport in the hallway.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|