Model Number DTPB2D4 |
Device Problems
Unable to Obtain Readings (1516); Capturing Problem (2891); Missing Test Results (3267)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) exhibited an invalid battery voltage with displayed questions marks for the longevity measurement.The device remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Correction: upon secondary review the reported event of invalid battery voltage with displayed questions marks for the longevity measurement did not cause or contribute to death or serious injury and is notlikely to do so if it were to recur.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) exhibited an invalid battery voltage with displayed questions marks for the longevity measurement.The device remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the battery measurement was not available.Analysis of the device memory indicated an issue with missing/invalid diagnostic data.Analysis of the device memory indicated an issue with capture management.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) was stuck in a non-wireless telemetry session.It was also noted that the data related to the device feature which automatically monitors pacing thresholds and adjust outputs data was not present on the report.The device was interrogated, and the persistent session active flag was cleared using a non-bluetooth telemetry session.The device remains in use.No patient complications have been reported as a result of this event.
|
|