• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 150 - AMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 150 - AMER Back to Search Results
Model Number 27003
Device Problem Degraded (1153)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/06/2022
Event Type  malfunction  
Event Description
It was reported to resmed that an astral device displayed an internal battery degraded warning alarm.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
The device was returned to resmed and an evaluation confirmed the complaint.The internal battery was replaced to address this issue.The device was serviced and fully tested before it was returned to the customer.Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the reported complaint was due to battery controller software.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - AMER
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key13694729
MDR Text Key289254083
Report Number3007573469-2022-00143
Device Sequence Number1
Product Code NOU
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27003
Device Catalogue Number27003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/08/2022
Distributor Facility Aware Date02/07/2022
Device Age82 MO
Date Report to Manufacturer03/08/2022
Date Manufacturer Received02/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2015
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-