• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TURON SHOULDER; GLENOID, KEELED, 50MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TURON SHOULDER; GLENOID, KEELED, 50MM Back to Search Results
Model Number 520-01-150
Device Problems Naturally Worn (2988); Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 02/21/2022
Event Type  Injury  
Event Description
Revision surgery: due to posterior wear of the glenoid.
 
Manufacturer Narrative
The reason for this revision surgery was reported as bone wear.The previous surgery and the surgery detailed in this event occurred 8 years and 4 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to bone wear.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, prolonged overhead activities, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Type of Device
GLENOID, KEELED, 50MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key13713447
MDR Text Key286913777
Report Number1644408-2022-00305
Device Sequence Number1
Product Code KWS
UDI-Device Identifier00888912145084
UDI-Public(01)00888912145084
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2014
Device Model Number520-01-150
Device Catalogue Number520-01-150
Device Lot Number53990639
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/21/2022
Initial Date FDA Received03/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/17/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
520-00-000 LOT 878C1271; 520-50-118 LOT 934C1047
Patient Outcome(s) Required Intervention;
Patient Age80 YR
Patient SexMale
-
-