• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME Back to Search Results
Model Number N/A
Device Problems Loose or Intermittent Connection (1371); Malposition of Device (2616)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).(b)(6).Investigation complete.This is an initial final report submission.Review of the most recent repair record determined the motor had signs of corrosion, failed the voltage test, and the speed was unstable.The needle bearing was worn, an e-ring was missing, and the ball plunger was damaged.The motor, needle bearing, e-ring, and ball plunger were replaced which resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that there was an issue in assembling the blade.It was loose and unstable.Power would not reach the blade.At product evaluation investigation the motor speed was found to be unstable.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key13723768
MDR Text Key286994331
Report Number0001526350-2022-00217
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number62042429
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/09/2022
Initial Date FDA Received03/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/23/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-