Visual inspection was performed on the graft bolt and the interbody spacer.Significant wear was observed on the distal threads of the graft bolt; this wear corresponds to large gouging in the posterior caudal end of the interbody spacer and in the inferior side of the medial bolt hole entry.The damage to the medial bolt hole and the posterior caudal endplate of the interbody spacer indicates that the graft bolt was driven at a considerably shallower angle than is indicated.As a result, the distal end of the graft bolt likely lodged in the porous titanium of the interbody spacer causing graft bolt torsional failure.Review of the device history records did not identify any manufacturing or processing related irregularities.The implants were found to be properly manufactured and released in accordance with design specifications.Labeling review: "possible adverse effects: possible adverse effects include: initial or delayed loosening, bending, dislocation, and/or breakage of device components.".
|