Model Number CSE-P-125-JA |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/24/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
On (b)(6) 2022 visual evaluation of the returned bowl from the cell saver® elite set - 125ml confirmed blood in the inner core with a crack in the inner core base.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due the event of the past report.
|
|
Event Description
|
On (b)(6) 2022, haemonetics was notified of a bowl inner core leak during a procedure in (b)(6), utilizing the cell saver® elite® autotransfusion system and cell saver® elite set 125ml.There was no reported impact to patients' health.
|
|
Search Alerts/Recalls
|