Model Number CSR60
Device Problems Vibration (1674); Noise, Audible (3273); Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2022
Event Type  malfunction  
Event Description
It was reported from (b)(6) that the compact speed reducer device made an abnormal noise, experienced vibration and generated heat.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once the investigation has been completed, and if additional information should become available, a supplemental medwatch report will be submitted accordingly.Udi: (b)(4).
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: this device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition of the device generating heat was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the device was frozen/would not move and had a component damage.It was further determined that the device had scratches on the shaft and was non-rotational.It was further determined that the device failed pretest for visual assessment.The assignable root cause was determined to be due to component failure from wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPACT SPEED REDUCER, 60:1
Type of Device
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kara ditty-bovard
4500 riverside drive
palm beach gardens, FL 33410
6103142063
MDR Report Key13769197
MDR Text Key287200301
Report Number1045834-2022-00275
Device Sequence Number1
Product Code HBB
UDI-Device Identifier00845384001782
UDI-Public00845384001782
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K960630
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Followup
Report Date 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCSR60
Device Catalogue NumberCSR60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2022
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/15/2022
Supplement Dates Manufacturer Received03/17/2022
Supplement Dates FDA Received03/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/08/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-