Model Number 133650 |
Device Problems
Device Slipped (1584); Detachment of Device or Device Component (2907)
|
Patient Problems
Hemorrhage/Bleeding (1888); Foreign Body In Patient (2687); Unspecified Tissue Injury (4559)
|
Event Date 03/02/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, two hours after a coronary artery bypass grafting, grafting of great saphenous vein, there was a dramatic blood pressure drop occurred and re-opening surgery was required.The clip was not found and reclipping was required.There was unanticipated tissue loss and extended incision.Hospitalization was also extended.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, two hours after a coronary artery bypass grafting, grafting of great saphenous vein, there was a dramatic blood pressure drop occurred and re-opening surgery was required to tie the vessel.All the missing clips fell into patient¿s cavity.The patient also experienced excess bleeding, however, the amount of blood loss is unknown.The clips were was not found and reclipping was required.There was unanticipated tissue loss and extended incision.Hospitalization was also extended.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the clip did not hold tightly to the vessel and detached and a component disengaged from the device into the surgical cavity.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: ensure that the handles are squeezed together firmly as far as they will go.Failure to squeeze the handles completely can result in clip malformation and possible bleeding or leakage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, two hours after a coronary artery bypass grafting, grafting of great saphenous vein, there was a dramatic blood pressure drop occurred and re-opening surgery was required to tie the vessel.All the missing clips fell into patient¿s cavity.The patient also experienced excess bleeding, however, the amount of blood loss is unknown.The clips were was not found and reclipping was required.There was unanticipated tissue loss and extended incision.Hospitalization was also extended.There will be no additional surgery to retrieved the clips.
|
|