STERIS CANADA ULC RELIANCE 444 WASHER/DISINFECTOR Back to Search Results
Device Problem Mechanical Problem (1384)
Patient Problem Laceration(s) (1946)
Event Date 02/17/2022
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite to inspect the reliance 444 washer/disinfector and spoke to user facility personnel regarding the reported event.The technician learned that the employee was attempting to manually clear an obstruction from the door when their finger was pinched by the conveyer arm resulting in the reported event.The technician inspected the washer and found that the teflon rack loading sliding guide which is part of the loading rack assembly inside the washer was missing a screw and was misaligned.This caused the racks that sit on the loading rack assembly to shift during the wash cycle and jam the door.The washer was installed in 2005 making it approximately 17 years old and is not under steris service agreement for maintenance activities; the user facility is responsible for all maintenance activities.The reliance 444 operator manual states (1-1), "warning - personal injury hazard: risk of pinch point between door and upper panel.Do not push on top portion of doors; do not push on door when door is rising; do not push on door when door is jammed." the operator manual further states (1-1), "warning - personal injury hazard: if an obstruction is present in the wash chamber door and door is unable to raise, do not attempt to remove obstruction from under the door.Door cables may have slackened and door may close at high speed when obstruction is removed.Call a qualified service technician to safely remove obstruction." the technician replaced the screw to the loading rack assembly, tested the unit, confirmed it to be operating according to specification, and returned it to service.The technician counseled the customer on the safe operating protocols, specifically to call a qualified service technician to safely remove obstructions.No additional issues have been reported.
 
Event Description
The user facility reported an employee obtained an injury to their finger while operating their reliance 444 washer/disinfector.The employee self-applied a bandage and returned to work.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIANCE 444 WASHER/DISINFECTOR
Type of Device
WASHER/DISINFECTOR
Manufacturer (Section D)
STERIS CANADA ULC
490 boulevard armand-paris
quebec, G1C 8 A3
CA  G1C 8A3
Manufacturer (Section G)
STERIS CANADA ULC
490 boulevard armand-paris
quebec, G1C 8 A3
CA   G1C 8A3
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key13825702
MDR Text Key290952008
Report Number9680353-2022-00011
Device Sequence Number1
Product Code MEC
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/18/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-