Model Number CI-1500-01 |
Device Problems
Electrical /Electronic Property Problem (1198); Loose or Intermittent Connection (1371); Intermittent Communication Failure (4038)
|
Patient Problems
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/05/2022 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing loss of lock.Device testing could not be performed due to loss of lock.
|
|
Manufacturer Narrative
|
A device history review was completed and it was noted that the device had rework in the printed circuit board.The device was fully conforming after rework and passed all required verifications.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock could not be obtained at any spacing, however lock was obtained after the completion of the residual gas analysis test.The condition of the electrode prevented an electrical test from being performed.The device passed some of the electrical test performed.The device passed the mechanical tests performed.The resistance test revealed elevated resistance was measured across the kovar tab.The reported complaint of no lock was confirmed during the analysis of this device.Elevated resistance was measured across an electrical component(kovar tab), which is believed to be related to the loss of lock.An internal corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|