• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ ADVANTAGE IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ ADVANTAGE IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1500-01
Device Problems Electrical /Electronic Property Problem (1198); Loose or Intermittent Connection (1371); Intermittent Communication Failure (4038)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2022
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing loss of lock.Device testing could not be performed due to loss of lock.
 
Manufacturer Narrative
A device history review was completed and it was noted that the device had rework in the printed circuit board.The device was fully conforming after rework and passed all required verifications.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock could not be obtained at any spacing, however lock was obtained after the completion of the residual gas analysis test.The condition of the electrode prevented an electrical test from being performed.The device passed some of the electrical test performed.The device passed the mechanical tests performed.The resistance test revealed elevated resistance was measured across the kovar tab.The reported complaint of no lock was confirmed during the analysis of this device.Elevated resistance was measured across an electrical component(kovar tab), which is believed to be related to the loss of lock.An internal corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ ADVANTAGE IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
jennifer rhudy
28515 westinghouse place
valencia, CA 91355
MDR Report Key13831241
MDR Text Key293931142
Report Number3006556115-2022-00467
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016872495
UDI-Public(01)07630016872495(11)200110(17)211130
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/30/2021
Device Model NumberCI-1500-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/20/2022
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Not provided
Supplement Dates FDA Received05/04/2022
07/22/2022
08/18/2022
02/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age5 YR
Patient SexFemale
-
-