Model Number CLV-S45 |
Device Problem
Optical Problem (3001)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/18/2022 |
Event Type
malfunction
|
Event Description
|
The user facility reported that during preparation for use (patient was not under anesthesia), the visera xenon light source lamp didn¿t work.Another similar device was used for the procedure.There was no report of patient or user harm due to the event.
|
|
Manufacturer Narrative
|
The device was sent to the olympus service center for evaluation.The device inspection found that the main light would not light up when the device was powered on, confirming the reported lamp failure.The issue was resolved by replacing the lamp.Additionally, the igniter was replaced due to the age of the device.The investigation is ongoing; therefore, a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 7 years since the subject device was manufactured.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, the reported phenomenon seemed to have occurred due to lamp failure.However, the cause of the phenomenon could not be identified.Olympus will continue to monitor field performance for this device.
|
|