• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VISERA XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VISERA XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S45
Device Problem Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2022
Event Type  malfunction  
Event Description
The user facility reported that during preparation for use (patient was not under anesthesia), the visera xenon light source lamp didn¿t work.Another similar device was used for the procedure.There was no report of patient or user harm due to the event.
 
Manufacturer Narrative
The device was sent to the olympus service center for evaluation.The device inspection found that the main light would not light up when the device was powered on, confirming the reported lamp failure.The issue was resolved by replacing the lamp.Additionally, the igniter was replaced due to the age of the device.The investigation is ongoing; therefore, a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 7 years since the subject device was manufactured.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, the reported phenomenon seemed to have occurred due to lamp failure.However, the cause of the phenomenon could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13831520
MDR Text Key287553633
Report Number8010047-2022-04715
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170275777
UDI-Public04953170275777
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/18/2022
Initial Date FDA Received03/20/2022
Supplement Dates Manufacturer Received04/05/2022
Supplement Dates FDA Received05/04/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-