The logbook of the affected evita v600 with serial number (b)(6), photos, service report and e-mail correspondence were available for the investigation.Information on the circuit system and accessories used, the use of a closed suction system without expiration of the suction maneuver, and compliance with the requirements for electromagnetic compatibility according to the instructions for use in the patient's room and ct were not provided.Since the patient required resuscitation in the temporal context, further information was requested regarding the patient's health status after the reported event.This revealed that the patient made a complete recovery shortly after the event.Based on the results of the investigation on site and with the manufacturer, no malfunction of the internal battery or its performance could be identified.The ventilator was disconnected from ac power at 9:29 a.M.System time on the reported day of the event, according to log analysis.The ventilator issued a "low battery" alarm after 23 minutes of battery operation, which corresponded to battery performance (internal nimh battery) within specifications.At 9:54 a.M., the user reconnected to the line power supply.Contrary to what was reported, the device did not fail during battery operation.During this time, or in close temporal proximity, the alarm message "deconnection detected" was repeatedly issued and automatically reset a few seconds later in each case, after reconnection was detected.The software function "anti-air-shower" was activated on the ventilator by the user, whereby the flow was automatically reduced in the event of a detected disconnection until the detected reconnection.This specified device behavior is described in the user manual, but may have been misinterpreted by the user as a stop of ventilation.After mains power was restored, other ventilation-related alarms followed, e.G., "minute volume low," "airway pressure low," "no inspiratory flow detected," and "pressure limited," which were more likely due to the application or patient situation and did not represent a technical malfunction.A few minutes later, the user briefly switched to standby, then back to ventilation, disconnected the device from the mains power supply again, and finally ended therapy with the device at 10:01.For safety, the ventilator's software analyzes and verifies correct device function.If the performance of the device is limited by a functional or application fault, the user is alerted to the condition by appropriate acoustic alarms as well as a visual message in the display of the control unit.The alarm is triggered according to the alarm limits set on the device.In the event of a mains power failure or during transport, the ventilator is powered by the internal nimh battery to continue operation.Switchover to battery supply is indicated by an alarm.The specified battery runtime with a new and fully charged internal nimh battery during normal ventilation (without gs500) is approx.30 minutes.Depending on the battery capacity and battery voltage, additional alarm messages "battery low" and "battery discharged" are displayed as the battery life progresses.In the event of a complete power failure, the safety valve automatically opens against the environment, thus allowing the patient to breathe spontaneously.The ventilator's auxiliary audible alarm is immediately activated to alert the user to the equipment failure.This alarm is powered from an independent power source and lasts for at least 2 minutes.The dräger service technician has subsequently checked the device for correct device function according to manufacturer's specifications.Based on the results of the investigation in the field and in the logbook analysis, no deviation from the specified device function or device failure was detected.The results of this complaint investigation did not reveal any new or additional risks that have not yet been addressed in the ventilator risk management file.The number of similar cases due to the same cause is within the expected range of the respective risk assessment and is therefore accepted.
|