According to the reporter, the doctor reported problems in the functioning of the peritoneal catheters.It was stated that the tip or tail of the catheters moved from its place and out of the pelvic hollow, this did not allow good flow or remove all fluid from the peritoneal cavity.It was mentioned that there was no leak on one of the catheters (2113300312).There were no patient symptoms or complications associated with this event, no medical or surgical intervention was needed to prevent a permanent impairment of a function, and the event did not lead to or extend patient hospitalization.There was no reported patient injury.
|
Additional information: d10, g1, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.The evaluation found no potentially contributing factors.It was reported that there was an ingrowth or catheter migration issue and there was an insufficient flow issue.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|