Model Number A1602
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/11/2022
Event Type  malfunction  
Manufacturer Narrative
The event unit is anticipated to return to applied medical for evaluation.A follow-up report will be submitted upon completion of investigation.
 
Event Description
Procedure performed: unknown.Event description: les clips ne ferment pas /pas de maintien de la prise.Translation: clips do not close / no hold on the clamp.Additional information received from applied medical representative via email on 24-feb-2022: there was no patient injury.3 clips were tried on the patient.The issue was resolved by taking another lot.The clips are available for return.The single-used devices were not reused, they used other clips of the same lot.Type of intervention: change of device.Patient status: no patient injury.
 
Event Description
Procedure performed: unknown.Event description: les clips ne ferment pas /pas de maintien de la prise.Translation: clips do not close / no hold on the clamp.Additional information received from applied medical representative via email on 24-feb-2022: there was no patient injury.3 clips were tried on the patient.The issue was resolved by taking another lot.The clips are available for return.The single-used devices were not reused, they used other clips of the same lot.Type of intervention: change of device.Patient status: no patient injury.
 
Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Testing was performed on the event unit; however, the customer¿s experience of clips unable to close could not be confirmed or replicated.Applied medical has reviewed the details surrounding the event and related product and is unable to determine the cause of the reported event or confirm that a product malfunction occurred.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
A1602, 6MM L/L STEALTH CLIP 1/4 F 10/BX
Type of Device
CLIP, VASCULAR
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key13847079
MDR Text Key288852680
Report Number2027111-2022-00490
Device Sequence Number1
Product Code DSS
UDI-Device Identifier00607915115074
UDI-Public(01)00607915115074(17)260518(30)01(10)1418686
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K883909
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA1602
Device Catalogue Number101374001
Device Lot Number1418686
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/24/2022
Initial Date FDA Received03/22/2022
Supplement Dates Manufacturer Received02/24/2022
Supplement Dates FDA Received06/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-