MPRI ACTIVA Back to Search Results
Model Number 3389-28
Device Problems Break (1069); Material Deformation (2976); Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2022
Event Type  malfunction  
Event Description
It was reported that many times during lead implantation (seen in several cases), the lead bends at the contact points and at the stem as soon as the stylet is removed causing inaccuracy in placement of the lead and will affect the patient outcomes.Follow-up with the rep indicated that no damage was seen when the lead was taken out of the package.The stylet was not bent when the stylet was removed either.The lead was straight and not damaged.They initially made a tract with the dbs cannula stylet and passed the lead in the tract.The same lead was implanted as the bend was not anticipated and happened only after removal of the stylet.The lead bend was seen as soon as the stylet was removed before the burr hole cap was placed.There were no environmental/external/patient factors that may have led or contributed to the issue.They tried reinserting the lead again after removing stylet but the issue was not resolved.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received: it was reported that there was no surgical intervention as they placed the same lead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13866390
MDR Text Key288117152
Report Number2649622-2022-05418
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3389-28
Device Catalogue Number3389-28
Device Lot NumberVA2F2PL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/22/2022
Date Device Manufactured06/17/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
Patient Weight60 KG
-
-