H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was returned for evaluation.The guidewire was broken and the guidewire tip decoiled.The helix was pointing out from the catheter.The tube was damaged at two positions, at 85 cm and 95 cm from the tip.A dense occlusion in catheter was observed at 10 cm to 20 cm distance from the tip.The investigation is confirmed for reported issues.A clear root cause could not be identified but a broken guidewire represents a known inherent risk.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: d4 (expiry date: 11/2023), g3.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|