• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO EUPHORA RX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO EUPHORA RX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, Back to Search Results
Model Number EUP3015X
Device Problems Fracture (1260); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2022
Event Type  malfunction  
Event Description
During a procedure, an attempt was made to use one euphora rx ptca balloon catheter to treat a severely tortuous, severely calcified lesion exhibiting 99% stenosis located at the mid/ostium left main (lm) coronary artery/ramus.The device was inspected with no issues noted.Negative prep was performed with no issues noted.It was reported that the device detached, cracked, or fractured at the catheter.It was stated that the shaft of the device fractured within the guide catheter when trying to pull the device out of the guide.It was stated that the balloon was being used to trap a.014'' wire in the guide catheter.The patient is reported to be alive with no injury.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: resistance was encountered when advancing or withdrawing the device.Excessive force was not used during delivery.The device was being used to pin the guidewire within the guide catheter.The balloon was inflated within the guide catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis summary: the device returned with a detachment on the transition shaft.The transition shaft material was jagged and uneven at the detachment site.Proximal to the detachment site, the transition shaft was stretched.Distal to the detachment site, numerous kinks were evident along the transition shaft and distal shaft.Blood was visable in the inflation lumen.The balloon returned deflated with crystalised residue and blood visible in the balloon.The distal tip was stretched.No other damage evident to the remainder of the device.Additional information: a review of the procedural images returned.Images showed initial poor blood flow in the left coronary system.The patient had previously been treated with coronary artery by-pass (cabg) due to complex coronary artery disease.Balloon inflation was completed in the mid to proximal left main on multiple occasions with a waist evident in the mid balloon area, most likely due to calcification in the vessel.Multiple interventional and imaging devices were used during the procedure.The reported fracture or the root cause was not identified in the images provided.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EUPHORA RX
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY,
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key13868892
MDR Text Key290809727
Report Number9612164-2022-01124
Device Sequence Number1
Product Code LOX
UDI-Device Identifier00643169560154
UDI-Public00643169560154
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143480
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/12/2023
Device Model NumberEUP3015X
Device Catalogue NumberEUP3015X
Device Lot Number223301914
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/10/2022
Initial Date FDA Received03/23/2022
Supplement Dates Manufacturer Received04/07/2022
05/16/2022
05/25/2022
Supplement Dates FDA Received04/21/2022
05/24/2022
05/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age82 YR
Patient SexMale
Patient Weight100 KG
-
-