A case of right ventricular outflow tract (rvot) perforation was reported in a procedure where the nykanen rf wire kit was used on a patient with pulmonary atresia and patent ductus arteriosus.The rvot was highly angulated.The nykanen rf wire was delivered to the target puncture site using a 4fr judkins right catheter.The inadvertent perforation occurred when the physician was attempting to puncture the pulmonary artery.The procedure was aborted and extracorporeal membrane oxygenation was employed.Cardiac function showed improvement afterwards.Central-shunt surgery was planned for a later date.The physician indicated the guiding catheter was not flexible enough for the angulated rvot.There is no evidence to suggest that baylis medical device caused or contributed to the reported incident.However, as a baylis medical device was reported to be among the devices use in the procedure, baylis medical has decided to submit this report.
|
There is no evidence to suggest that the baylis medical device caused or contributed to the reported incident and there is no suspected device malfunction.However, as the baylis medical devices were reported to be among the devices used in the procedure, baylis medical has decided to submit this report.A dhr review has been completed for the lot in question and all the devices in the lot met relevant requirements prior to release.
|