Model Number G151
Device Problem Unexpected Therapeutic Results (1631)
Patient Problem Ventricular Fibrillation (2130)
Event Date 01/21/2022
Event Type  Injury  
Event Description
It was reported that the patient implanted with this cardiac resynchronization therapy defibrillator (crt-d) presented after receiving a shock.Review of the store episodes found anti-tachycardia pacing (atp) had been delivered along with three shocks.Atp was not effective in converting the arrhythmia and rhythm acceleration was noted after which a shock was delivered.The patient remained in a slow ventricular tachycardia (vt) below the detect rate.It was reported that the patient in on the heart transplant list and had been externally shocked three times since presenting.However, after successful conversion the patient goes back into a ventricular arrhythmia.Amiodarone was administer and further consultation with cardiology was planned.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13910629
MDR Text Key288034970
Report Number2124215-2022-04512
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534638
UDI-Public00802526534638
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/29/2017
Device Model NumberG151
Device Catalogue NumberG151
Device Lot Number111121
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/21/2022
Initial Date FDA Received03/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/29/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Hospitalization;
Patient Age63 YR
Patient SexMale
-
-