Model Number FT4 G3
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.
 
Event Description
There was an allegation of questionable elecsys ft4 iii and elecsys ft3 iii results for 1 patient sample on a cobas 8000 e 602 module.The module's serial number was requested, but not provided.This medwatch covers the alleged results for the ft4 assay.Please refer to medwatch (b)(4) for the alleged ft3 assay results.The results were not reported outside of the laboratory.
 
Manufacturer Narrative
The customer's calibration and qc results and sample pre-analytic details were requested but not provided.From the provided information, a general reagent and interfering factor issue could be excluded.The observed differences in ft4 values generated with the roche assay and the abbott assay are caused by differences in the overall setups of the assays, the antibodies used, differences in reference materials, and the differences in the standardization methodology used.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key13926783
MDR Text Key297947589
Report Number1823260-2022-00877
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976887190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/03/2022
Initial Date FDA Received03/28/2022
Supplement Dates Manufacturer Received04/07/2022
Supplement Dates FDA Received04/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-