STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP SILVERGLIDE PROSERIES BIPOLAR FORCEPS, BAYONET, 24CM / 2.0MM TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Model Number 6730240020 |
Device Problem
Flaked (1246)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during testing at the user facility sterile processing department, the device insulation was found to be chipped, posing the risk of a material being lost in a surgical site.There were no adverse consequences related to this event.
|
|
Manufacturer Narrative
|
Follow-up report submitted to document device evaluation results.H3 other text : device discarded at user facility.
|
|
Event Description
|
It was reported that during testing at the user facility sterile processing department, the device insulation was found to be chipped, posing the risk of a material being lost in a surgical site.There were no adverse consequences related to this event.
|
|