STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP SILVERGLIDE BIPOLAR PRECISION FORCEPS BAYONET, 22CM / 1.2MM TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Model Number 6720220012 |
Device Problem
Flaked (1246)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/21/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during testing in the sterile processing department of the user facility, insulation was peeling off near the tips of the device, posing the risk of a small component being lost in a surgical site.There were no adverse consequences related to this event.
|
|
Manufacturer Narrative
|
Follow-up report submitted to document device evaluation results.Correction: b5: during service evaluation at the manufacturer facility, the technician found the device insulation was actually intact, and it was device tip plating material that was flaking off and missing.
|
|
Event Description
|
It was reported that during testing in the sterile processing department of the user facility, insulation was peeling off near the tips of the device, posing the risk of a small component being lost in a surgical site.There were no adverse consequences related to this event.During service evaluation at the manufacturer facility, the technician found the device insulation was actually intact, and it was the device tip plating material that was flaking off and missing.
|
|