ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME, COMPLETE Back to Search Results
Model Number N/A
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2021
Event Type  malfunction  
Manufacturer Narrative
This complaint has been recorded under (b)(4).Review of the most recent repair record determined the device was out of calibration at the 0 reading and the width plates were damaged.The device was recalibrated and 1-4 inch width plates were replaced.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the item was sent in for repair for an unknown reason.Upon inspection it was found that the width plates were damaged on the device.No harm or delay were reported.No other adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME, COMPLETE
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key13930344
MDR Text Key288057348
Report Number0001526350-2022-00279
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100000
Device Lot Number31067200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2021
Was the Report Sent to FDA? No
Date Manufacturer Received03/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/05/1996
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-