In addition to describe event or problem, the suspect device was returned to olympus for evaluation.Upon inspection, the loop was visualized at the tip of the coil sheath.The loop had detached from the hook, but both the hook and loop were caught in the coil sheath.There were no physical abnormalities or deformation visualized that could have led to the reported malfunction.The investigation is ongoing; therefore, a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
|
The customer reported that during an endoscopic polypectomy procedure, using a single use ligating device, the polyp was looped for ligation.After ligation, the loop seemed to interfere with the sheath and would not come off from the polyp.The details of how the loop was removed were unknown.There was indication that the loop may have been torn off with force; however, there was no report that surgical intervention was required.There was no report of patient injury due to the event.
|
The aware date should be 26-jan-2022.Investigation activities have been opened to manage the actions related to this report and any required mdr reporting.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been almost 1 year since the subject device was manufactured.Based on the results of the investigation, the investigation confirmed the loop was detached from the hook and was caught in the coil sheath.The specific root cause of the loop being detached from the hook and caught in the coil sheath could not be determined at this time.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1.The loop was detached from the hook while the coil sheath was being retracted into the tube sheath, or the loop was hung on to the tissue and it was positionally fixed at the distal end of the tube.As a result, the loop was unable to release.2.The tube sheath was pulled toward the proximate side while the hook was extended from the distal end of the coil sheath.3.Since the tube sheath was pulled toward the proximate side, causing the tube to pull the loop.This caused the loop to retract into the coil sheath.As a result, the loop was caught in between the hook and inner surface of the coil, and the loop stopped moving.The event can be prevented by following the instructions for use which state: ¿do not strike or crush the coil sheath during operation.Doing so can damage the distal end of the coil sheath, which could make it impossible to detach the loop after ligation.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual.¿ ¿do not remove the loop from the hook while the coil sheath is not extended from the tube sheath.Otherwise, the loop may be tangled with the hook and become impossible to be removed.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual.¿ ¿do not hold the loop with the distal end of the tube sheath while the loop is surrounding the tissue.Otherwise, when the tissue is ligated, the loop may be detached from the hook in the tube sheath and tangled with the hook.That may make the loop impossible to be removed.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual.¿ ¿never use excessive force to operate the instrument.This could damage the instrument.¿ olympus will continue to monitor field performance for this device.
|