• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON GMBH BD¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD; CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON GMBH BD¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD; CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL Back to Search Results
Catalog Number 257306
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2022
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported while testing with bd columbia cna agar with 5% sheep blood contaminated plates were discovered.There was no indication that results were reported out, and there was no report of patient impact.The following information was provided by the initial reporter: contaminated plates.
 
Manufacturer Narrative
H6: investigation summary it was reported that some plates would have show contamination.The complaint trends were reviewed and similar complaints were found for this lot number.Therefore, a mrb (material review board) was performed and this issue will be investigated within a capa (corrective and preventive action) 4344556.Batch history record (bhr) review: the batch history review did not indicate any discrepancies.All release testing was satisfactory, and no deviations were observed.Sample analysis: picture was provided showing the reported contamination.The retention samples were reviewed and the plates were found to be without deviation.Evaluation results: at this stage of our investigation, we have excluded any systemic failure in our manufacturing process.Please note that this product does not have an sal claim.It is filled aseptically, therefore an occurrence of a contamination event cannot be entirely ruled out.We only release product batches to the market with an aql (acceptable quality level) = 1.5.Although this contamination level is very low, we cannot completely exclude that a customer may receive a contaminated plate.As several complaints were reported for this product lot number regarding contamination, a mrb (mrb 22-006) and capa (# 4344556) were created for further evaluation.The root cause investigation is currently ongoing within capa 4344556.Based upon our investigation the complaint was confirmed.H3 other text : see h10.
 
Event Description
It was reported while testing with bd columbia cna agar with 5% sheep blood contaminated plates were discovered.There was no indication that results were reported out, and there was no report of patient impact.The following information was provided by the initial reporter: contaminated plates.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD
Type of Device
CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL
Manufacturer (Section D)
BECTON DICKINSON GMBH
8-12 tullastrasse
heidelberg 69126
GM  69126
Manufacturer (Section G)
BECTON DICKINSON GMBH
8-12 tullastrasse
heidelberg 69126
GM   69126
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13934702
MDR Text Key288692557
Report Number9680577-2022-00041
Device Sequence Number1
Product Code JSJ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/28/2022
Device Catalogue Number257306
Device Lot Number2033567
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-