• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Model Number 261221
Device Problem Fail-Safe Did Not Operate (4046)
Patient Problem Brain Injury (2219)
Event Date 03/14/2022
Event Type  Injury  
Event Description
A physician reported disengagement failure of a perforator during surgery causing dural damage.Dural suture was performed.The perforator was used with elan4 (aesculap).The procedure was completed with a replacement product.The event led to less than 30 minutes surgical delay.It is unknown if the perforator clicked in place with the drill or if the recommended spring tests were performed between each burr hole.
 
Manufacturer Narrative
The device involved in the reported incident is not available for evaluation.An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Manufacturer Narrative
Lot number provided: dhr - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Udi - (b)(4).Manufacturing date is december 17, 2021.Expiration date is november 30, 2026.
 
Manufacturer Narrative
Updated fields:  d4, d9, g3, g6, h2, h3, h10.Udi #: (b)(4).The disposable perforator (id 261221) was not returned for evaluation (as per customer, product not available) and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and device history record (dhr) could not be reviewed.The root cause(s) of the reported issue could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event Description
N/a.
 
Event Description
N/a.
 
Manufacturer Narrative
Perforator was returned for evaluation: failure analysis - the perforator unit was inspected using the unaided eye.Anomalies observed were a heavily soiled perforator with organic matter, showing signs of rust.Ifu testing was performed, and the following was observed; the unit had a frozen inner/outer drill which had to be freed.Once free, the unit performed as intended.Functional testing was performed using the same protocol it underwent at finished goods testing prior to release.The unit was found to perform as intended and fulfilled the acceptance criteria.The complaint could not be verified through failure analysis.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.Product was received for analysis and the investigation could not confirm the complaint.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA
Manufacturer (Section G)
RAYNHAM
325 paramount drive
raynham MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13939613
MDR Text Key288118209
Report Number3014334038-2022-00066
Device Sequence Number1
Product Code HBF
UDI-Device Identifier10381780513599
UDI-Public10381780513599
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number261221
Device Catalogue Number261221
Device Lot Number5958506
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/15/2022
Initial Date FDA Received03/29/2022
Supplement Dates Manufacturer Received03/30/2022
04/19/2022
06/29/2022
Supplement Dates FDA Received04/12/2022
04/19/2022
07/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/17/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient SexMale
-
-