CELESTICA DE MONTERREY S.A. DE C.V. SURGINEEDLE; PNEUMOPERITONEUM NEEDLE Back to Search Results
Model Number 172015
Device Problems No Device Output (1435); Retraction Problem (1536)
Patient Problems Hemorrhage/Bleeding (1888); Unspecified Tissue Injury (4559)
Event Date 03/04/2022
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic procedure, at the beginning, the safety device did not work when the device was inserted; the needle did not retract and was exposed inside the abdominal cavity, damaging the patient's liver.A new device was used to resolve the issue.
 
Event Description
According to the reporter, during a laparoscopic cholecystectomy, at the beginning, the safety device did not work when the device was inserted; the needle did not retract and was exposed inside the abdominal cavity, damaging the patient's liver.There was a limited liver bleeding.To resolve that, the surgeon coagulated the tissue with no consequences.A new device was used to resolve the issue.
 
Manufacturer Narrative
Additional info: b5, g3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic cholecystectomy, at the beginning, the safety device did not work when the device was inserted; the needle did not retract and was exposed inside the abdominal cavity, damaging the patient's liver.It was noted that the protector spring lost sensitivity as if it had not triggered, there was no "click" when it was entered.There was a limited liver bleeding.To resolve that, the surgeon coagulated the tissue with no consequences.A new device was used to resolve the issue.
 
Manufacturer Narrative
Correction: b5, g3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: b2, g1 (manufacturer name, mfr contact first name, last name, street 1, mfr city, region, postal code, email, phone number), g3 correction: b5 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic cholecystectomy, at the beginning, the safety device did not work when the device was inserted; the needle did not retract and was exposed inside the abdominal cavity, damaging the patient's liver.It was noted that the protector spring lost sensitivity as if it had not triggered, and there was no "click" when it was entered.There was a limited liver bleeding.The issue happened with two devices.To resolve that, the surgeon coagulated the tissue with no consequences.A new device was used to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGINEEDLE
Type of Device
PNEUMOPERITONEUM NEEDLE
Manufacturer (Section D)
CELESTICA DE MONTERREY S.A. DE C.V.
calle octava #102
parque industrial monterrey
apodaca nuevo leon
MX 
Manufacturer (Section G)
CELESTICA DE MONTERREY S.A. DE C.V.
calle octava #102
parque industrial monterrey
apodaca nuevo leon
MX  
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key13939808
MDR Text Key288120216
Report Number3003139373-2022-00089
Device Sequence Number1
Product Code FHO
UDI-Device Identifier10884521080805
UDI-Public10884521080805
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K863330
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number172015
Device Catalogue Number172015
Device Lot NumberT1K287X
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2022
Initial Date FDA Received03/29/2022
Supplement Dates Manufacturer Received05/02/2022
05/26/2022
09/07/2022
Supplement Dates FDA Received05/17/2022
06/10/2022
09/23/2022
Date Device Manufactured10/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient SexMale
-
-