The device and the lens were returned separately in the opened blister tray inside the opened carton.The lens stop and plunger lock were removed.The plunger was oriented correctly.Viscoelastic was observed in the device.Blood was observed dried at the nozzle tip.The plunger was retracted to mid-nozzle.The lens was returned inside a folded piece of white gauze.Viscoelastic and blood were dried on the lens.The optic was cut or torn into two portions.One optic portion was torn and gouged on the anterior surface.Product history records were reviewed and documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The root cause for the reported complaint cannot be determined.The used device and the damaged lens were returned separately.The optic was torn and gouged, interpreted as the complaint.The plunger was retracted.The plunger position in relation to the lens during advancement cannot be determined.The instructions for use (ifu) instructs: after the lens has been advanced to the nozzle line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.After confirming the lens is properly positioned and the haptics are folded properly, proceed with lens implantation.Proceeding with implantation of a misfolded haptic or a lens that appears to be ¿out of position¿ can result in a broken haptic or other negative outcome, since the haptic may be trapped and stretched, and/or pinched and sheared by the moving plunger.A qualified viscoelastic was indicated.Plunger override may occur: due to rapid advancement faster than the ifu recommend rate.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device allowing the plunger to override the lens.If the device is overfilled with ovd, this can prevent the trailing haptic from being placed properly or move the lens out of position resulting in misfolding.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.The manufacturer internal reference number is: (b)(4).
|
A physician reported that during a cataract surgery with intraocular lens (iol) implant procedure, when the surgeon tried to insert the lens after setting with ophthalmic viscosurgical devices (ovd), it was confirmed that the behavior of the iol was unstable.The plunger passed over the optical part and the iol optical part had a crack.The iol was cut in half, removed it out of the eye and the surgery was completed after replacing the another backup one.
|